RadiCS application class Body region/methods Maximum display luminance cd/m2 Maximum luminance ratio Matrix of the image rendering device Luminance characteristic curve Homogeneity of the image rendering device Homogeneity between multiple devices
I. Mammography >=250 >=250:1 >=2048 x >=2048 GSDF +/- 10% <25 % <10 %
II. Mammographic stereotaxis >=200 >=100:1 >=1024 x >=1024 GSDF +/- 15% <25 % <20 %
III. Projection radiography (thorax, skeleton, abdomen) >=250 >=250:1 >=1600 x >=1200 GSDF +/- 10% <25 % <20 %
IV. Fluoroscopy, all applications >= 150 >= 100:1 >=1024 x >=1024 GSDF +/- 15% <25 % <20 %
V. Computer tomography >= 150 >= 100:1 >=1024 x >=1024 GSDF +/- 15% <25 % <20 %
VI. For RK 5: Dental digital volume tomography intraoral X-ray diagnostics with dental X-ray tube heads, panoramic tomographs, teleradiograph exposures of the cranium, dental cranial overview exposures, hand radiographs to detect skeletal growth >= 200 >= 100:1 >=1024 x >=768 - <30 % <30 %
VII. For RK 6: Intraoral X-ray diagnostics with dental X-ray tube heads, panoramic tomographs, teleradiograph exposures of the cranium, dental cranial overview exposures, hand radiographs to detect skeletal growth >= 300 >= 100:1 >=1024 x >=768 - <30 % <30 %
VIII. Clinical review - - - GSDF - -
Updated: 2016/02, Subject to changes